论文标题

氯丁橡胶,蜂蜡和Carbomer-Gel的声学和超声特征,以模拟超声

Acoustic and Ultrasonographic Characterization of Neoprene, Beeswax, and Carbomer-Gel to Mimic Soft-tissue for Ultrasound

论文作者

Phani, Debjani, V., Rajasekhar K., Thomas, Anjali, Paramu, Raghukumar, Singh, M. Suheshkumar, S., Shaiju V., Muraleedharan, Venugopal, Nair, R. K.

论文摘要

与软组织类似的声学量的材料与模仿组织超声(US)的材料必不可少。声学数量由声速(CUS),声阻抗(AI)和衰减系数组成。在这项工作中,确定了氯丁橡胶橡胶,蜂蜡和碳纤维 - 凝胶的声量。使用脉冲回波技术估算了CUS和衰减系数。 AI是根据密度和CU的乘积计算得出的。将结果与基于国际辐射单位和测量委员会的基准进行了比较。接受标准为1.043 g/cm3(密度),1561 m/s(cus),1.63 mrayls(AI)和衰减系数在0.5-0.7 dB/cm/mhz之内。获得了计算机断层扫描(CT)和US的标本图像,以与人肝(临床软组织)的各个图像进行比较,以评估图像对比度和回声性的相似性。氯丁橡胶和蜂蜡的结果不令人满意。但是,Carbomer-Gel的声量(密度1.03 g/cm3,CUS 1567 m/s和AI 1.61 mrayls和AI 1.61 mrayls和衰减系数0.6 db/cm/mHz)在2%之内是令人满意的。 Carbomer-Gel图像可以有效地模仿肝图像的对比度和回声。 Neoderene-50,蜂蜡和C-凝胶的CUS测量不确定性分别为0.36%,0.11%和0.28%。这些样品中的衰减系数的不确定性为4.2%,1.9%和2.6%。 Carbomer-Gel的结果可能对我们来说类似于软组织。它含有95%的水,毫不费力地准备,并且可以支持开发低成本的幻影,以定期评估美国扫描仪并为患者护理做出贡献。

Materials with acoustic quantities similar to soft-tissue are essential as tissue-mimicking materials for diagnostic ultrasound (US). Acoustic quantity consists of the sound velocity (cus), acoustic impedance (AI) and attenuation coefficient. In this work, the acoustic quantities of neoprene rubber, beeswax, and Carbomer-gel were determined. The cus and attenuation coefficient were estimated using the pulse-echo technique. The AI was calculated from the product of density and cus. Results were compared with a benchmark based on the International Commission on Radiation Units and Measurements Report-61, Tissue Substitutes, Phantoms and Computational Modelling in Medical Ultrasound. The acceptance criteria were 1.043 g/cm3 (density), 1561 m/s (cus), 1.63 MRayls (AI) and attenuation coefficients within 0.5-0.7 dB/cm/MHz. Computerized tomography (CT) and US images of specimens were obtained to compare with respective images of the human liver (a clinical soft-tissue), to evaluate the similarities in image contrast and echogenicity. Results of neoprene and beeswax were unsatisfactory.However, the acoustic quantities of Carbomer-gel (density 1.03 g/cm3, cus 1567 m/s, and AI 1.61 MRayls, and attenuation coefficient 0.6 dB/cm/MHz) were satisfactory within 2%. Carbomer-gel images could efficiently mimic the contrast and echogenicity of liver images. The uncertainties in cus measurements were 0.36 %, 0.11 % and 0.28 % for neoprene-50, beeswax and C-gel respectively. The attenuation coefficients had uncertainties 4.2 %, 1.9 %, and 2.6 % in these samples. The results of Carbomer-gel could resemble soft-tissue for US. It contains 95 % water, is effortless to prepare, and can support in developing a low-cost phantom for periodic performance evaluation of US scanners and contribute to patient care.

扫码加入交流群

加入微信交流群

微信交流群二维码

扫码加入学术交流群,获取更多资源